Troubleshooting

This section provides troubleshooting information in the form of an easy to read table. If a problem occurs during the first operation of a new fryer, recheck Installation and Maintenance procedures.

To isolate a malfunction, proceed as follows:

  • Clearly define the problem (or symptom) and when it occurs.

  • Locate the problem in the Troubleshooting table.

  • Review all possible causes. Then, one-at-a-time, work through the list of corrections until the problem is solved.

Problem Cause Solution
Product Color Not Correct: A. Too Dark (some batches)
  • Temperature programmed too hot.

  • Breading product too far in advance

  • Done alarm ignored for more than 20 seconds.

  • Wrong product button pressed.

  • See Diagnostic Mode D 10; if temperature settings have been changed, have the controls reintialized.

  • Bread product just before frying.

  • If the fryer hasn’t been used since the problem batch, see Information Mode 4 H; for more information on this problem, see Information Modes 5 U, 6 U, 7 R, or 8 R

  • Be sure to press the correct product button; if the fryer hasn’t been used since the problem batch, see Information Mode 4 B to see what product button was pressed

B. Too Dark (all batches)
  • Temperature probe out of calibration.

  • Peanut oil too old.

  • Peanut oil too dark.

  • Faulty probe “E6”.

  • See Diagnostic Mode D 1 to adjust color of product.

  • Check temperature probe calibration; see Checking Temperature Probe Calibration Section; if less than 15 degrees off, have probe calibrated; if more than 15 degrees off, replace probe.

  • If peanut oil is smoking or has burnt taste, change peanut oil.

  • See Diagnostic Mode D 2; Change peanut oil if controls indicate it should be changed.

  • Filter peanut oil or change peanut oil.

  • If probe can’t be recalibrated, have probe replaced.

C. Too Light (all batches)
  • Temperature probe out of calibration.

  • Slow fryer heat-up/recovery.

  • Oil usage wasn’t set for new for

  • peanut oil

  • See Diagnostic Mode D 1 to adjust color of product. Check temperature probe calibration; see Checking Temperature Probe Calibration Section; if less than 15 degrees off, have probe calibrated; if more than 15 degrees off, replace probe.

  • See Diagnostic Mode D 4 for present day’s performance; or see Information Modes 5, 6, 7, 8, and 9 for more information on this problem.

  • Low voltage; see Diagnostic Mode D 3 for present day’s voltage performance; see Information Modes 4, 5, 6, 7, 8, 9, and 15 for more information on this problem.

  • Low gas pressure; have gas pressure checked going to burners, on gas fryers.

  • See Diagnostic Mode D 2 for the age of the oil; see section 3-7 for setting the age of the oil.

D. Too Light (some batches)
  • Temperature programmed too low.

  • Product placed in peanut oil before proper temperature

  • Wrong product button pushed

  • Cook Cycle aborted before alarm and “DONE” flashes.

  • Frozen product placed in peanut oil.

  • See Diagnostic Mode D 10; if temperature settings have been changed, without authorization, have the controls reintialized.

  • If fryer hasn’t been used since the problem batch, see Information Mode 4 C; for more information on this see Information Modes 5 S, 6 S, 7 P, or 8 P.

  • If fryer hasn’t been used since problem batch, see Information Mode 4 B to see what product was selected.

  • See Diagnostic Mode D 7 (Continued) to see how many times the Cook Cycle was stopped before the end of the cycle.

  • Use fresh or thawed product; see Diagnostic Mode D 5 to see if the controls sensed any frozen or overloaded batches.

Dryness of Product
  • Moisture loss prior to cooking

  • Over-cooking the product

  • Time of Cook Cycle set too long

  • Wrong product button pushed

  • Low operating pressure

  • Use fresh product.

  • Cover product with plastic wrap, reducing evaporation.

  • Done alarm ignored for more than 20 seconds; if the fryer hasn’t been used since the problem batch, see Information Mode 4 H; for more information on this problem, see Information Modes 5 U, 6 U, 7R, or 8 R.

  • See Diagnostic Mode D 10; if time settings have been changed, have the controls reintialized.

  • If fryer hasn’t been used since problem batch, see Information Mode 4 B to see what product was selected.

  • Check pressure gauge reading. Check for pressure leaks.

Burned Taste

  • Burned peanut oil flavor.

  • Peanut oil needs filtering.

  • Vat not properly cleaned.

  • Replace peanut oil.

  • Filter peanut oil more often.

  • Drain and clean vat.

Product not done
  • Cook Cycle aborted before alarm, and “DONE” flashes

  • Frozen product placed in peanut oil

  • Wrong product button pushed

  • Temperature programmed too low or not programmed properly

  • Temperature probe out of calibration

  • Slow fryer heat-up/recovery

  • Product too thick

  • See Diagnostic Mode D 7 to see how many times the Cook Cycle was stopped before the end of the cycle.

  • Use fresh or thawed product; see Diagnostic Mode D 5 to see if the controls sensed frozen or overloaded batches.

  • If fryer hasn’t been used since problem batch, see Information Mode 4 B to see what product was selected.

  • See Diagnostic Mode D 10; if temperature settings have been changed, have the controls reintialized.

  • Check temperature probe calibration; see Checking Temperature Probe Calibration Section; a. If less than 5° off, see Diagnostic Mode D 1 b. If between 5 and 15 degrees off, calibrate probe; if more than 15° off, replace probe.

  • See Diagnostic Mode D 4 for present day’s performance; or see Information Modes 5, 6, 7, 8, and 9 for more information on this problem.

  • Low voltage; see Diagnostic Mode D 3 for present day’s voltage performance; see Information Modes 4, 5, 6, 7, 8, 9, and 15 for more information on this problem.

  • Low gas pressure; have gas pressure checked going to burners, on gas fryers.

  • Make sure chicken filets have been fileted.

With COOK/PUMP Switch in COOK position, fryer is completely without power Open circuit
  • Check to see if fryer is plugged in.

  • Check wall circuit breaker or fuse.

  • Have a qualified service technician check power supply and COOK/PUMP switch.

Pressure will not exhaust at end of valve cooking cycle.
  • Exhaust line from solenoid valve to exhaust tank clogged.

  • Solenoid valve clogged.

  • Deadweight clogged.

  • Exhaust line to stack clogged.

  • Turn unit off and allow fryer to cool to release pressure from vat; have all pressure lines,- exhaust stacks, and exhaust tank cleaned.

  • Have solenoid checked and

  • cleaned.

  • Turn unit off and allow fryer to cool to release pressure from vat; remove deadweight and clean, per Cleaning the Deadweight Valve.

  • Clean exhaust line to stack.

    To avoid severe personal injury, do not operate unit if pressure gauge shows high pressure conditions.

    Immediately place the cook/pump switch in the off position, which releases the pressure by allowing the unit to cool. Do not resume use of unit until cause of high pressure has been found and corrected.

Pressure does not build.

 

Not enough product in vat. Place full capacity product in vat when using fresh oil.
Metal shipping spacer not removed from deadweight. Remove shipping spacer; see Unpacking.
Lid open or not latched. Close and latch lid.
Solenoid valve leaking or not closing. Have solenoid valve checked or cleaned.
Deadweight assembly leaking. Have deadweight assembly repaired.
Pressure not programmed. See Diagnostic Mode D 10; if pressure settings have been changed, have the controls reintialized.

Lid gasket leaking.

Reverse or replace lid gasket; see Reversing the Lid Gasket.

Safety relief valve leaking. Check and replace if necessary.

Filter motor runs but pumps peanut oil slowly

  • Pump clogged.

  • Filter line connection loose.

  • Solidified peanut oil in lines.

  • Have pump cleaned

  • Tighten all filter line connections.

  • Clear all filter lines of solidified peanut oil.

Filter switch on, motor does not run
  • Defective COOK/PUMP switch

  • Defective motor

  • Motor thermal protector tripped

  • Have switch checked

  • Have motor checked

  • Reset thermal protector

  • per Resetting the Filter Pump Motor.

Motor hums but will not pump Clogged lines or pump
  • Have pump and lines removed and cleaned.

  • Have pump seal, rotor and rollers replaced.

Peanut oil will not heat

Blown fuse or tripped circuit breaker.

Reset breaker or replace fuse.

Faulty cord and plug

Check cord and plug.

Faulty PC board

Have control panel checked.

Faulty or tripped high limit “E10”

Reset high limit per Operating Components; if high limit doesn’t reset, have it checked.

Drain valve open “E15”

Close drain valve.

Possible faulty probe “E6”

Have temperature probe checked.

Possible faulty contactor (electric models)

See Diagnostic Modes D 4; see if “CHECK COILS, CONTACTORS AND WIRING” displays.

Gas valve knob turned to the OFF position (gas models)

Make sure the gas valve knob is turned to ON.

Faulty thermocouple on gas control valve (gas models)

Have thermocouple checked.

  • Faulty COOK/PUMP switch

  • Faulty drain switch “E15”

  • Possible faulty gas control

See Information Mode10 and check to see if the input code is present; if not, have fryer checked by a certified service technician.

Peanut oil heating slowly

 

  • Low or improper amps

  • Low or improper voltage

  • See Infomation Mode 16 for present amperage; or see Information Modes 4, 5, 6, 7, 8, 9, for more information on this problem; Diagnostic Mode D 4 gives present day’s heating performance.

  • See Diagnostic Modes D 3 & D 4 for present day’s voltage and heating performance; or see Information Modes 4, 5, 6, 7, 8, 9, and 15 for more information on this problem.

  • Weak or burnt out elements (elec. model)

  • Burnt or charred connectors

  • Faulty contactor (electric models)

  • Wire(s) loose

  • See Diagnostic Modes D 4; elements (elec. model) see if “CHECK COILS, CONTACTORS AND WIRING” shows on display; if so, have fryer checked by a certified service technician

  • Have wires tightened.

  • Supply line too small - low gas volume (gas models)

  • Improper ventilation system (gas models)

  • Increase supply line size; refer to installation instructions.

  • Refer to installation instructions.